“Data exchange” service offers individual users metadata transfer in several different formats. Citation formats are offered for transfers in texts as for the transfer into internet pages. Citation formats include permanent links that guarantee access to cited sources. For use are commonly structured metadata schemes : Dublin Core xml and ETUB-MS xml, local adaptation of international ETD-MS scheme intended for use in academic documents.
Export
Kolarević, Goran
Dozimetrijska verifikacija kliničkog sistema za planiranje radioterapije
Dozvoljavate umnožavanje, distribuciju i javno saopštavanje dela, i prerade, ako se navede ime autora na način odredjen od strane autora ili davaoca licence. Ova licenca ne dozvoljava komercijalnu upotrebu dela. Osnovni opis Licence: http://creativecommons.org/licenses/by-nc/3.0/rs/deed.sr_LATN Sadržaj ugovora u celini: http://creativecommons.org/licenses/by-nc/3.0/rs/legalcode.sr-Latn
Academic metadata
Doktorska disertacija
Medicinske nauke
Univerzitet u Banjoj Luci
Medicinski fakultet
Other Theses Metadata
Dosimetric verification of clinical radiotherapy treatment planning system
biomedicinske nauke - biofizika - medicinska fizika / biomedical sciences - byophysics - medical physics
Mirjanić, Dragoljub (mentor)
Moderna radioterapija (RT) danas predstavlja tehničko-tehnološki najsloženiju
medicinsku disciplinu, gde se jonizujuće zračenje koristi kao lek u lečenju malignih bolesti. Uz pomoć savremene tehnologije, zračenje se precizno aplikuje na volumen tkiva u kome se nalaze tumorske ćelije sa ciljem trajnog uništenja, dok se nastoji da se okolno zdravo tkivo maksimalno poštedi.
Ova dozimetrijska studija se bazira na tzv. End-to-End testu, koji pomoću heterogenog fantoma Thorax 002LFC, dimenziono i anatomski simulirajući ljudski toraks, kontroliše tačnost svih bitnih karika u RT lancu.
Osnovni ciljevi ove dozimetrijske studije su bili istražiti da li postoji značajna razlika u: a) dozimetrijskoj kalkulaciji sistema za planiranje radioterapije (TPS) u odnosu na vrednosti dobijene merenjem u fantomu na medicinskom linearnom akceleratoru (Linak) za 3D konformalnu (3DCRT), intenzitetom modulisanu (IMRT) i zapreminski modulisanu lučnu (VMAT) radioterapijsku tehniku zračenja, b) tačnosti dozimetrijskog proračuna između kalkulacionih algoritama Analytical Anisotropic Algorithm (AAA) i Acuros XB u različitim tkivima i energijama fotonskih snopova (6 MV i 16 MV), c) konverzionim krivama dobijenim CIRS 062M pelvis i Thorax 002LFC fantomima, pri
istim kondicijama kompjuterizovane tomografije (CT) - simulatora. Navedeni Thorax fantom je skeniran na CT simulatoru, na osnovu čega su napravljeni
RT planovi u TPS-u, te je ozračen i izmerene su apsolutne doze na Linaku.
Provedenim istraživanjem dobijena je potvrda ispravnog računanja doze TPS-a u
odnosu na pacijentu Linakom isporučenu. Dokazano je da postoji značajna razlika
relativnih grešaka između fotonskih snopova, tj. TPS preciznije računa RT tretmane sa 6 MV u odnosu na 16 MV. Utvrđena je statistički značajno manja relativna greška kod AAA vs Acuros XB, za kost. CT konverzione krive zavise od napona u CT cevi, a fantom Thorax 002LFC može se koristiti za kontrolu CT konverzione krive (kao deo kontrole kvaliteta, End-to-End testa).
U kliničkoj praksi pacijenti na CT simulatoru trebaju se snimati samo na naponu CT cevi na kom je urađena konverziona kriva (Haunsfildove jedinice (HU) u relativnu elektronsku gustinu (RED)), uneta u TPS. Trebalo bi razmotriti mogućnosti da RT TPS, za jedan CT simulator ima opciju sa više konverzionih krivih, za različite napone CTcevi, prečnika CT rekonstruisanog vidnog polja (FOV) i dimenzije lokalizacije koja se snima.
Modern radiotherapy (RT) today is technically and technologically the most complex medical discipline where ionizing radiation is used as a medicine in the treatment of malignant diseases. With the help of modern technology, radiation is precisely applied to the volume of tissue in which tumor cells are located with the aim of permanent destruction, while striving to save the surrounding healthy tissue as much as possible. This dosimetric study is based on the so-called End-to-End test, which uses the heterogeneous phantom Thorax 002LFC, which dimensionally and anatomically simulates the human thorax, to control the accuracy of all-important links in the RT chain.
The main objectives of this dosimetric study were to investigate whether there is a significant difference in: a) dosimetric calculation of radiotherapy planning system (TPS) in relation to the values obtained by measuring in a phantom on a medical linear accelerator (Linak) for 3D conformal (3DCRT), intensity modulated (IMRT) and volume modulated arc (VMAT) radiotherapy technique, b) accuracy of dosimetric calculation between calculation algorithms Analytical Anisotropic Algorithm (AAA) and Acuros XB in different tissues and energies of photon beams (6 MV and 16 MV), c) conversion curve obtained by CIRS 062M pelvis and Thorax 002LFC phantoms, at the same conditions of computed tomography (CT) - simulator. We scanned the Thorax phantom on a CT simulator, on the basis of which RT plans were made in TPS and irradiated it and measured the absolute doses on Linak. The conducted research confirmed the correct TPS calculation of the dose in relation to the patient delivered with Linak. We have proved that there is a significant difference in relative errors between photon beams, ie TPS more accurately calculates RT treatments with 6 MV compared to 16 MV. We found a statistically significantly smaller relative error in AAA vs Acuros XB, for bone. CT conversion curves depend on the CT tube voltage and the phantom Thorax 002LFC can be used to control the CT conversion curve (as part of the quality control, End-to-End test).In clinical practice, patients on a CT simulator should be imaged only at the voltage of the CT tube on which the conversion curve was done, Hounsfield units (HU) to relative electron density (RED) and entered into the TPS. Possibilities should be considered that
the RT TPS, for one CT simulator, has an option with multiple conversion curves, for different CT tube voltages, the diameter of the reconstructed field of view (FOV) and the dimensions of the localization being recorded.
CT simulator, CT kalibraciona kriva, heterogeni fantom, TPS, End - to - End test, kalkulacioni algoritmi
CT simulator, CT conversion curve, heterogeneous phantom, TPS, End - to - End test, calculation algorithms
prema CERIF šifrarniku : B - 140
Serbian
Moderna radioterapija (RT) danas predstavlja tehničko-tehnološki najsloženiju
medicinsku disciplinu, gde se jonizujuće zračenje koristi kao lek u lečenju malignih bolesti. Uz pomoć savremene tehnologije, zračenje se precizno aplikuje na volumen tkiva u kome se nalaze tumorske ćelije sa ciljem trajnog uništenja, dok se nastoji da se okolno zdravo tkivo maksimalno poštedi.
Ova dozimetrijska studija se bazira na tzv. End-to-End testu, koji pomoću heterogenog fantoma Thorax 002LFC, dimenziono i anatomski simulirajući ljudski toraks, kontroliše tačnost svih bitnih karika u RT lancu.
Osnovni ciljevi ove dozimetrijske studije su bili istražiti da li postoji značajna razlika u: a) dozimetrijskoj kalkulaciji sistema za planiranje radioterapije (TPS) u odnosu na vrednosti dobijene merenjem u fantomu na medicinskom linearnom akceleratoru (Linak) za 3D konformalnu (3DCRT), intenzitetom modulisanu (IMRT) i zapreminski modulisanu lučnu (VMAT) radioterapijsku tehniku zračenja, b) tačnosti dozimetrijskog proračuna između kalkulacionih algoritama Analytical Anisotropic Algorithm (AAA) i Acuros XB u različitim tkivima i energijama fotonskih snopova (6 MV i 16 MV), c) konverzionim krivama dobijenim CIRS 062M pelvis i Thorax 002LFC fantomima, pri
istim kondicijama kompjuterizovane tomografije (CT) - simulatora. Navedeni Thorax fantom je skeniran na CT simulatoru, na osnovu čega su napravljeni
RT planovi u TPS-u, te je ozračen i izmerene su apsolutne doze na Linaku.
Provedenim istraživanjem dobijena je potvrda ispravnog računanja doze TPS-a u
odnosu na pacijentu Linakom isporučenu. Dokazano je da postoji značajna razlika
relativnih grešaka između fotonskih snopova, tj. TPS preciznije računa RT tretmane sa 6 MV u odnosu na 16 MV. Utvrđena je statistički značajno manja relativna greška kod AAA vs Acuros XB, za kost. CT konverzione krive zavise od napona u CT cevi, a fantom Thorax 002LFC može se koristiti za kontrolu CT konverzione krive (kao deo kontrole kvaliteta, End-to-End testa).
U kliničkoj praksi pacijenti na CT simulatoru trebaju se snimati samo na naponu CT cevi na kom je urađena konverziona kriva (Haunsfildove jedinice (HU) u relativnu elektronsku gustinu (RED)), uneta u TPS. Trebalo bi razmotriti mogućnosti da RT TPS, za jedan CT simulator ima opciju sa više konverzionih krivih, za različite napone CTcevi, prečnika CT rekonstruisanog vidnog polja (FOV) i dimenzije lokalizacije koja se snima.